Modifiable Settings
Default Settings Table 11, “Default Settings” displays the Haemonetics default settings for the
Cell Saver Elite device:
Table 11, Default Settings
Cell Salvage
Parameter Values Default
Fat Reduction On/Off Off
Auto-Fill On/Off On
Auto-Wash On/Off/Ask/
Skip
On
Partial Bowl Wash Ask/Single/
Double
Ask
Pump Regulation On/Off On
a
Empty Destination Blue/Red Blue
Volume Accounting
b
Reservoir/
Circuit
Reservoir
Final Cycle On/Off Off
Smart Empty On/Off On
Parameter 225 mL Bowl 125 mL Bowl 70 mL Bowl Min Max Step
Fill Start Volume (mL) 800 800 400 200 3000 100
Fill Resume Volume (mL) 400 400 200 200 3000 100
Fill Pump Rate
c
(mL/min) 500 225 100 25 1000 25
Minimum Wash Volume
(mL)
1000 750 300 500
d
5000 250
e
Wash Pump Rate
c
(mL/
min)
450 200 100 25 1000 25
Empty Pump Rate
f
(mL/
min)
500/400/100 300/150/100 100 25 1000 25
Emergency Fill Pump
Rate (mL/min)
800 400 N/A 25 1000 25
Emergency Wash Pump
Rate (mL/min)
800 400 N/A 25 1000 25
Emergency Empty Pump
Rate
f
(mL/min)
500/400/250 300/250/100 N/A 25 1000 25