Chapter 1, Introduction 19
Haemonetics
®
Cell Saver
®
Elite
®
User Manual P/N 120745-IE, Manual revision: AA
Device Specifications
Note: The use of materials not provided or recommended by Haemonetics is
the sole responsibility of the end-user, and the end-user will be responsible for
any personal injury and/or property damage related to such use.
Device
Classification
The Cell Saver Elite is classified as a continuous operation, Class I, Type CF,
IPX1 device, as defined by IEC/EN 60601 standards for medical electrical
equipment.
Physical
Specifications
The approximate dimensions and weight of the Cell Saver Elite device are as
follows:
The noise level of the Cell Saver Elite device is < 70 dB.
Environmental
Specifications
The following environmental conditions should be respected pertaining to
operation and storage of the Cell Saver Elite device.
Warning: Equipment not suitable for use in the presence of a flammable
anesthetic mixture with air or with oxygen or nitrous oxide.
Table 3, Physical Specifications
Depth/cm (in.) Height/cm (in.) Width/cm (in.)
Device Alone 54.6 cm
(21.5 in)
41.9 cm
(16.5 in)
29.8 cm
(11.75 in)
Device With Cart
IV poles extended 67.3 cm
(26.5 in)
182.9 cm
(72 in)
53.3 cm
(21 in)
IV poles down 67.3 cm
(26.5 in)
121.9 cm
(48 in)
53.3 cm
(21 in)
Weight of device 25 kg (56 lbs)
Weight of cart 18 kg (39 lbs)