90 Chapter 5, General Operation: Cell Salvage
P/N 120745-IE, Manual revision: AA Haemonetics
®
Cell Saver
®
Elite
®
User Manual
1. Touch (Pause).
2. Touch End Procedure. A confirmation screen appears.
3. Touch End Procedure and wait for the device to empty the bowl (if full)
and purge the blue line.
4. Once the blue line has been emptied completely, remove the current
processing set from the device.
5. Install a new processing set, following the instructions beginning on page
81.
6. Once the new processing set is installed, touch Resume Procedure.
All procedure statistics from the procedure will be retained, and suction can
remain on throughout this process.
Changing the
Bowl Size
During a
Procedure
If the user selected the incorrect bowl size from the Bowl Selection screen,
navigate to the Records screen, view the procedure record, and edit the
processing set on the Disposables tab, following the instructions on page 136.