Suction
Specifications
The specifications for the Cell Saver Elite suction are as follows.
Laser
Specifications
The Cell Saver Elite device is a class 3R laser product.
The laser specifications for the Cell Saver Elite device are as follows:
Table 7, Suction Specifications
Characteristics Values
SmartSuction
Recommended reservoir volume 3 L
Recommended A&A line length 12 ft [3.6 m]
Recommended A&A line inner
diameter
0.3 in [7.6 mm]
Recommended suction tip inner
diameter
0.3 in [7.6 mm]
Operating vacuum 20 to 150 mmHg
(2.7 to 20.0 kPa; 26.7 to 200 mbar)
Vacuum cutoff 175 mmHg
(23.3 kPa; 233 mbar)
Maximum free air flow 40 L/min
Manual Suction
Operating vacuum 50 to 250 mmHg
(6.7 to 33.3 kPa; 66.7 to 333.3
mbar)
Maximum free air flow 40 L/min
Post-Op Suction
Operating vacuum 25 to100 mmHg
(3.3 to 13.3 kPa; 33.3 to 133.3
mbar)
Maximum free air flow 40 L/min
Table 8, Laser Specifications
Characteristics Values
Max radiation output 3 mW
Wavelength 650 nm
Max light output 7 mW (bowl optics)
1.7 mW +/- 0.2 mW (barcode reader)
Standards IEC/EN 60825-1:2007
a