EasyManuals Logo

Haemonetics Cell Saver Elite User Manual

Haemonetics Cell Saver Elite
220 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #21 background imageLoading...
Page #21 background image
Chapter 1, Introduction 21
Haemonetics
®
Cell Saver
®
Elite
®
User Manual P/N 120745-IE, Manual revision: AA
Suction
Specifications
The specifications for the Cell Saver Elite suction are as follows.
Laser
Specifications
The Cell Saver Elite device is a class 3R laser product.
The laser specifications for the Cell Saver Elite device are as follows:
Table 7, Suction Specifications
Characteristics Values
SmartSuction
Recommended reservoir volume 3 L
Recommended A&A line length 12 ft [3.6 m]
Recommended A&A line inner
diameter
0.3 in [7.6 mm]
Recommended suction tip inner
diameter
0.3 in [7.6 mm]
Operating vacuum 20 to 150 mmHg
(2.7 to 20.0 kPa; 26.7 to 200 mbar)
Vacuum cutoff 175 mmHg
(23.3 kPa; 233 mbar)
Maximum free air flow 40 L/min
Manual Suction
Operating vacuum 50 to 250 mmHg
(6.7 to 33.3 kPa; 66.7 to 333.3
mbar)
Maximum free air flow 40 L/min
Post-Op Suction
Operating vacuum 25 to100 mmHg
(3.3 to 13.3 kPa; 33.3 to 133.3
mbar)
Maximum free air flow 40 L/min
Table 8, Laser Specifications
Characteristics Values
Max radiation output 3 mW
Wavelength 650 nm
Max light output 7 mW (bowl optics)
1.7 mW +/- 0.2 mW (barcode reader)
Standards IEC/EN 60825-1:2007
a

Table of Contents

Other manuals for Haemonetics Cell Saver Elite

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Haemonetics Cell Saver Elite and is the answer not in the manual?

Haemonetics Cell Saver Elite Specifications

General IconGeneral
BrandHaemonetics
ModelCell Saver Elite
CategoryLaboratory Equipment
LanguageEnglish

Related product manuals