162 Chapter 11, Troubleshooting
P/N 120745-IE, Manual revision: AA Haemonetics
®
Cell Saver
®
Elite
®
User Manual
Event Messages
The following section provides a list of event messages, which appear on the
display screen when an action is required by the user.
Table 16, Event Messages
ID # Event Message Text
102 Effluent Line Sensor Failure
Explanation:
An issue has been detected with the effluent line sensor. The procedure can be performed, but
the device cannot monitor the effluent from the bowl.
Corrective Action:
To perform the procedure with the user monitoring the effluent:
1. Touch Continue.
2. Monitor the effluent from the bowl during processing.
If red cell spillage occurs, manually decrease the pump speed.
OR
To end the procedure:
1. Touch End Procedure.
2. If in power-on self-test, power down the device.
If the problem persists, service is required to resolve this issue.
103 Manifold Pressure Sensor Failure
Explanation:
An issue has been detected with the manifold pressure sensor.
Corrective Action:
Restart the device.
If the problem persists, service is required.

Table of Contents

Other manuals for Haemonetics Cell Saver Elite

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Haemonetics Cell Saver Elite and is the answer not in the manual?

Haemonetics Cell Saver Elite Specifications

General IconGeneral
BrandHaemonetics
ModelCell Saver Elite
CategoryLaboratory Equipment
LanguageEnglish

Related product manuals