Chapter 11, Troubleshooting 209
Haemonetics
®
Cell Saver
®
Elite
®
User Manual P/N 120745-IE, Manual revision: AA
407 System Fault
Explanation:
An issue has been detected with the system.
Corrective Action:
Restart the device.
If the problem persists, service is required.
408 Software Update in Progress
Explanation:
A software update is in progress.
Corrective Action:
Wait for all steps to complete and then restart the device.
409 Software Update Error
Explanation:
The software update was not completed successfully.
Corrective Action:
Retry the software update.
If the problem persists, service is required.
500 Confirmation Required
Explanation:
The software update option was selected. This will install the selected software version and will
permanently erase all user data including procedure records and settings. The software update
process can take several minutes to complete and will require restarting the device when
complete. Confirmation is required to proceed.
Corrective Action:
Touch Update Software to continue the software update process.
OR
Touch Cancel to cancel the software update process.
Table 16, Event Messages