Chapter 5, General Operation: Cell Salvage 83
Haemonetics
®
Cell Saver
®
Elite
®
User Manual P/N 120745-IE, Manual revision: AA
Figure 38, Inserting the bowl into the centrifuge chuck
4. 70 mL bowl only: Ensure that the red indicator lines inside the chuck
adaptor are visible.
5. Position the header arm around the top of the bowl.
6. Turn the latch on the header arm clockwise until it locks into place.
7. Spin the bowl to ensure it spins freely.
Figure 39, Closing and locking the header arm
Note: A click will be heard when the locking mechanism is completely secured.
Warning: Verify that the outlet port and effluent tubing are free of any
restrictions prior to initiating a procedure. If the outlet port is inadvertently
clamped off, the bowl rotary seal may become compromised. See
“Avoiding
Flow Restrictions” on page 67
for more information.
Installing the Effluent Line Sensor Tubing
To install the effluent line tubing:
1. Floss the effluent tubing into the effluent line sensor groove.
2. Ensure that the tubing is completely installed in the effluent line sensor.
1.
2.
3.
4.
1. Header arm
2. Tubing in effluent line
sensor
3. Header arm latch
4. Bowl in centrifuge chuck