100Nickel LR
Add 1 mL of HI 93740B 0.3% PAN
solution to each beaker, cap and
swirl to mix.
Press TIMER and the display will show a countdown prior to
adding reagent C or, alternatively, wait for 15 minutes. Then, in
both cases, add one packet of HI 93740C EDTA reagent to each
beaker, cap and swirl to mix until complete dissolution.
Fill one cuvet up to the mark with 10 mL of
the blank.
Place the cuvet into the holder and
ensure that the notch on the cap is
positioned securely into the groove.
Press ZERO and "SIP" will appear on the display.
Wait for a few seconds and the display will
show "-0.0-". Now the meter is zeroed and
ready for measurement.
Fill a second cuvet up to the mark with 10 mL
of the reacted sample.
Insert the second cuvet into the in-
strument.
10 mL
10 mL