MANGANESE HIGH RANGEMANGANESE HIGH RANGE
MANGANESE HIGH RANGEMANGANESE HIGH RANGE
MANGANESE HIGH RANGE
10 mL
SPECIFICATIONS
Range 0.0 to 20.0 mg/L
Resolution 0.1 mg/L
Accuracy ±0.2 mg/L ±3% of reading
Typical EMC ±0.1 mg/L
Deviation
Light Source Tungsten lamp with narrow band interference
filter @ 525 nm
Method Adaptation of the Standard Methods for the
Examination of Water and Wastewater, 18
th
edition, Periodate method. The reaction between
manganese and reagents causes a pink tint in
the sample.
REQUIRED REAGENTS
Code Description Quantity
HI 93709A-0 Citrate 1 packet
HI 93709B-0 Sodium periodate 1 packet
REAGENT SETS
HI 93709-01 Reagents for 100 tests
HI 93709-03 Reagents for 300 tests
For other accessories see page 141.
MEASUREMENT PROCEDURE
Select the program number corresponding to Manganese HR on
the secondary LCD by pressing PROGRAM W and V.
Fill the cuvet with 10 mL of unreacted sample (up
to the mark) and replace the cap.
Place the cuvet into the holder and
ensure that the notch on the cap is
positioned securely into the groove.
Press ZERO and “SIP” will blink on the display.
Manganese HR

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Hanna Instruments C 99 series and is the answer not in the manual?

Hanna Instruments C 99 series Specifications

General IconGeneral
BrandHanna Instruments
ModelC 99 series
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals