Nitrite HR
NITRITE HIGH RANGENITRITE HIGH RANGE
NITRITE HIGH RANGENITRITE HIGH RANGE
NITRITE HIGH RANGE
SPECIFICATIONS
Range 0 to 150 mg/L
Resolution 1 mg/L
Accuracy ±4 mg/L ±4% of reading
Typical EMC ±1 mg/L
Deviation
Light Source Tungsten lamp with narrow band interference
filter @ 575 nm
Method Adaptation of the Ferrous Sulfate method. The
reaction between nitrite and the reagent causes a
greenish-brown tint in the sample.
REQUIRED REAGENTS
Code Description Quantity
HI 93708-0 Powder reagent 1 packet
REAGENT SETS
HI 93708-01 Reagents for 100 tests
HI 93708-03 Reagents for 300 tests
For other accessories see page 141.
MEASUREMENT PROCEDURE
Select the program number corresponding to Nitrite HR on the
secondary LCD by pressing PROGRAM W and V.
Fill the cuvet up to the mark with 10 mL
of unreacted sample and replace the cap.
Place the cuvet into the holder and
ensure that the notch on the cap is
positioned securely into the groove.
Press ZERO and “SIP” will blink on the display.
Wait for a few seconds and the display will
show “-0.0-”. Now the meter is zeroed
and ready for measurement.
10 mL

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Hanna Instruments C 99 series and is the answer not in the manual?

Hanna Instruments C 99 series Specifications

General IconGeneral
BrandHanna Instruments
ModelC 99 series
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals