29
BROMINEBROMINE
BROMINEBROMINE
BROMINE
Bromine
SPECIFICATIONS
Range 0.00 to 8.00 mg/L
Resolution 0.01 mg/L
Accuracy ±0.08 mg/L ±3% of reading
Typical EMC ±0.01 mg/L
Deviation
Light Source Tungsten lamp with narrow band interference
filter @ 525 nm
Method Adaptation of the Standard Methods for the
Examination of Water and Wastewater, 18
th
edition, DPD method. The reaction between bro-
mine and the reagent causes a pink tint in the
sample.
REQUIRED REAGENTS
Code Description Quantity
HI 93716-0 DPD Reagent 1 packet
REAGENT SETS
HI 93716-01 Reagents for 100 tests
HI 93716-03 Reagents for 300 tests
For other accessories see page 141.
MEASUREMENT PROCEDURE
• Select the program number corresponding to Bromine on the
secondary LCD by pressing PROGRAM W and V.
• Fill the cuvet with 10 mL of unreacted sample (up
to the mark) and replace the cap.
• Place the cuvet into the holder and
ensure that the notch on the cap is
positioned securely into the groove.
• Press ZERO and “SIP” will blink on the display.
10 mL