EasyManuals Logo

Hanna Instruments C 99 series User Manual

Hanna Instruments C 99 series
148 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #56 background imageLoading...
Page #56 background image
56
CYANIDECYANIDE
CYANIDECYANIDE
CYANIDE
Cyanide
SPECIFICATIONS
Range 0.000 to 0.200 mg/L
Resolution 0.001 mg/L
Accuracy ±0.005 mg/L ±3% of reading
Typical EMC Dev. ±0.001 mg/L
Light Source Tungsten lamp with narrow band interference
filter @ 610 nm
Method Adaptation of the Standard Methods for the
Examination of Water and Wastewater, 18
th
edition, Pyridine-Pyrazolone method. The reac-
tion between cyanide and reagents causes a
blue tint in the sample.
REQUIRED REAGENTS
Code Description Quantity
HI 93714A-0 Reagent A 1 spoon
HI 93714B-0 Reagent B 1 packet
HI 93714C-0 Reagent C 1 packet
REAGENT SETS
HI 93714-01 Reagents for 100 tests
HI 93714-03 Reagents for 300 tests
For other accessories see page 141.
MEASUREMENT PROCEDURE
Select the program number corresponding to Cyanide on the
secondary LCD by pressing PROGRAM W and V.
Fill the cuvet with 10 mL of unreacted sample (up
to the mark) and replace the cap.
Place the cuvet into the holder and
ensure that the notch on the cap is
positioned securely into the groove.
Press ZERO and “SIP” will blink on the display.
Wait for a few seconds and the display will
show “-0.0-”. Now the meter is zeroed
and ready for measurement.
10 mL

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Hanna Instruments C 99 series and is the answer not in the manual?

Hanna Instruments C 99 series Specifications

General IconGeneral
BrandHanna Instruments
ModelC 99 series
CategoryMeasuring Instruments
LanguageEnglish