102Nitrate
NITRATENITRATE
NITRATENITRATE
NITRATE
SPECIFICATIONS
Range 0.0 to 30.0 mg/L
Resolution 0.1 mg/L
Accuracy ±0.5 mg/L ±10% of reading
Typical EMC ±0.1 mg/L
Deviation
Light Source Tungsten lamp with narrow band interference
filter @ 525 nm
Method Adaptation of the cadmium reduction method.
The reaction between nitrate and the reagent
causes an amber tint in the sample.
REQUIRED REAGENTS
Code Description Quantity
HI 93728-0 Powder reagent 1 packet
REAGENT SETS
HI 93728-01 Reagents for 100 tests
HI 93728-03 Reagents for 300 tests
For other accessories see page 141.
MEASUREMENT PROCEDURE
• Select the program number corresponding to Nitrate on
the secondary LCD by pressing PROGRAM W and V.
• Using the pipette, fill the cuvet with 6 ml of sample,
up to half of its height, and replace the cap.
• Place the cuvet into the holder and
ensure that the notch on the cap is
positioned securely into the groove.
• Press ZERO and “SIP”
will blink on the dis-
play.
• Wait for a few seconds and the display will
show “-0.0-”. Now the meter is zeroed
and ready for measurement.
6 mL