124
PHOSPHATE LOW RANGEPHOSPHATE LOW RANGE
PHOSPHATE LOW RANGEPHOSPHATE LOW RANGE
PHOSPHATE LOW RANGE
Phosphate LR
SPECIFICATIONS
Range 0.00 to 2.50 mg/L
Resolution 0.01 mg/L
Accuracy ±0.04 mg/L ±4% of reading
Typical EMC ±0.01 mg/L
Deviation
Light Source Tungsten lamp with narrow band interference
filter @ 610 nm
Method Adaptation of the Ascorbic Acid method. The
reaction between phosphate and the reagent
causes a blue tint in the sample.
REQUIRED REAGENTS
Code Description Quantity
HI 93713-0 Powder reagent 1 packet
REAGENT SETS
HI 93713-01 Reagents for 100 tests
HI 93713-03 Reagents for 300 tests
For other accessories see page 141.
MEASUREMENT PROCEDURE
• Select the program number corresponding to Phosphate LR on
the secondary LCD by pressing PROGRAM W and V.
• Rinse, cap and shake the cuvet several times
with unreacted sample. Fill the cuvet with
10 mL of sample up to the mark and replace
the cap.
• Place the cuvet into the holder and
ensure that the notch on the cap is
positioned securely into the groove.
• Press ZERO and “SIP” will blink on the display.
10 mL