SPECIFICATIONS
Range 0.00 to 10.00 mg/L
Resolution 0.01 mg/L
Accuracy ±0.05 mg/L ±5% of reading
Typical EMC ±0.01 mg/L
Deviation
Light Source Tungsten lamp with narrow band interference
filter @ 420 nm
Method Adaptation of the ASTM Manual of Water and
Environmental Technology, D1426-92, Nessler
method. The reaction between ammonia and
reagents causes a yellow tint in the sample.
REQUIRED REAGENTS
Code Description Quantity
HI 93715A-0 First Reagent 4 drops (6 drops in seawater)
HI 93715B-0 Second Reagent 4 drops (10 drops in seawater)
REAGENT SETS
HI 93715-01 Reagents for 100 tests
HI 93715-03 Reagents for 300 tests
For other accessories see page 141.
MEASUREMENT PROCEDURE
• Select the program number corresponding to Ammonia MR on the
secondary LCD by pressing PROGRAM W and V.
• Fill the cuvet with 10 mL of unreacted sample (up
to the mark) and replace the cap.
• Place the cuvet into the holder and
ensure that the notch on the cap is
positioned securely into the groove.
• Press ZERO and “SIP” will blink on the display.
• Wait for a few seconds and the display will
show “-0.0-”. Now the meter is zeroed
and ready for measurement.
Ammonia MR
10 mL