SPECIFICATIONS
Range 0.000 to 1.000 mg/L
Resolution 0.001 mg/L
Accuracy ±0.005 mg/L ±10% of reading
Typical EMC ± 0.001 mg/L
Deviation
Light Source Tungsten lamp with narrow band interference
filter @ 575 nm.
Method Adaptation of the PAN method. The reaction
between silver and reagents causes an orange
tint in the sample.
REQUIRED REAGENTS
Code Description Quantity
HI 93737A-0 Buffer Reagent A 1 mL
HI 93737B-0 Buffer Reagent B 1 mL
HI 93737C-0 Indicator Reagent C 2 mL
HI 93737D-0 Fixing Reagent D 2 mL
HI 93703-51 Dispersing Agent 4-6 drops
(only when necessary, see note)
REAGENT SETS
HI 93737-01 Reagents for 50 tests
HI 93737-03 Reagents for 150 tests
For other accessories see page 141.
MEASUREMENT PROCEDURE
Select the program number corresponding to Silver on the
secondary LCD by pressing PROGRAM W and V.
Note: for best results perform your tests between 20-24°C.
Fill two graduated beakers with 25 mL of sample.
Add 1.0 mL of HI 93737A reagent
to only one beaker (the blank) and
swirl gently to mix.
SILVERSILVER
SILVERSILVER
SILVER
Silver
# 1 # 2
# 1

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Hanna Instruments C 99 series and is the answer not in the manual?

Hanna Instruments C 99 series Specifications

General IconGeneral
BrandHanna Instruments
ModelC 99 series
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals