130Silica
INTERFERENCES
Interference may be caused by:
Phosphate above 60 mg/L (causes a 2% reduction in reading)
Phosphate above 75 mg/L (causes an 11% reduction in reading)
Sulfide and high concentration of iron
Eliminate color and turbidity interferences by zeroing the meter with
the original water sample.