FLUORIDEFLUORIDE
FLUORIDEFLUORIDE
FLUORIDE
Fluoride
SPECIFICATIONS
Range 0.00 to 2.00 mg/L
Resolution 0.01 mg/L
Accuracy ±5% of reading
Typical EMC ±0.01 mg/L
Deviation
Light Source Tungsten lamp with narrow band interference
filter @ 575 nm
Method Adaptation of the Standard Methods for the
Examination of Water and Wastewater, 18
th
edition, SPADNS method. The reaction between
fluoride and the liquid reagent causes a red tint
in the sample.
REQUIRED REAGENT
Code Description Quantity
HI 93729-0 SPADNS Reagent 4 mL
REAGENT SETS
HI 93729-01 Reagents for 100 tests
HI 93729-03 Reagents for 300 tests
For other accessories see page 141.
MEASUREMENT PROCEDURE
• Select the program number corresponding to Fluoride on the
secondary LCD by pressing PROGRAM W and V.
• Add 2 mL of HI 93729 SPADNS
reagent to two cuvets.
• Fill one of the cuvets with distilled water
(up to the mark), replace the cap and
invert several times to mix.
• Fill the other cuvet with sample (up to
the mark), replace the cap and invert
several times to mix.
10 ml
10 ml
# 1
# 2
2 ml