EasyManua.ls Logo

Hanna Instruments C 99 series - Zinc Measurement; Zinc Specifications; Zinc Required Reagent; Zinc Reagent Sets

Hanna Instruments C 99 series
148 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
134
ZINCZINC
ZINCZINC
ZINC
Zinc
SPECIFICATIONS
Range 0.00 to 3.00 mg/L
Resolution 0.01 mg/L
Accuracy ±0.03 mg/L ±3% of reading
Typical EMC ±0.01 mg/L
Deviation
Light Source Tungsten lamp with narrow band interference
filter @ 575 nm
Method Adaptation of the Standard Methods for the
Examination of Water and Wastewater, 18
th
edition, Zincon method. The reaction between zinc
and the reagents causes an orange to a dark
violet tint in the sample.
REQUIRED REAGENT
Code Description Quantity
HI 93731A-0 Reagent A 1 packet
HI 93731B-0 Reagent B 0.5 mL
REAGENT SETS
HI 93731-01 Reagents for 100 tests
HI 93731-03 Reagents for 300 tests
For other accessories see page 141.
MEASUREMENT PROCEDURE
Select the program number corresponding to Zinc on the secondary
LCD by pressing PROGRAM W and V.
Fill one graduated mixing cylinder up to the
20 mL mark with the sample.
Add the content of one packet of HI 93731A zinc reagent, close
the cylinder and invert several times to mix until dissolution is
complete.
20 mL

Table of Contents

Related product manuals