EasyManuals Logo

Hanna Instruments C 99 series User Manual

Hanna Instruments C 99 series
148 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #79 background imageLoading...
Page #79 background image
79
HYDRAZINEHYDRAZINE
HYDRAZINEHYDRAZINE
HYDRAZINE
Hydrazine
SPECIFICATIONS
Range 0 to 400 µg/L
Resolution 1 µg/L
Accuracy ±4% of full scale
Typical EMC ±2 µg/L
Deviation
Light Source Tungsten lamp with narrow band interference
filter @ 420 nm
Method Adaptation of the ASTM Manual of Water and
Environmental Technology, method D1385-88,
p-Dimethylaminobenzaldehyde method. The reac-
tion between hydrazine and the liquid reagent
causes a yellow tint in the sample.
REQUIRED REAGENT
Code Description Quantity
HI 93704-0 Liquid Reagent 24 drops
REAGENT SETS
HI 93704-01 Reagents for 100 tests
HI 93704-03 Reagents for 300 tests
For other accessories see page 141.
MEASUREMENT PROCEDURE
Select the program number corresponding to Hydrazine on the
secondary LCD by pressing PROGRAM W and V.
Fill one cuvet up to the mark with 10
mL of distilled water.
Place the cap and insert the cuvet
# 1 into the holder. Ensure that
the notch on the cap is positioned
securely into the groove.
Press ZERO and SIP will blink while the
instrument is adjusting the light level.
10 mL
# 1
# 1

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Hanna Instruments C 99 series and is the answer not in the manual?

Hanna Instruments C 99 series Specifications

General IconGeneral
BrandHanna Instruments
ModelC 99 series
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals