80Hydrazine
Wait for a few seconds and the display will
show a blinking "good". Now the meter is
ready to take a zero measurement.
Fill a second cuvet with 10 mL of unreacted
sample (up to the mark) and replace the cap.
Add 12 drops of the
HI 93704 reagent
to each cuvet. Re-
place the caps and
shake gently to mix.
Place the blank (#1) into the holder
and ensure that the notch on the cap
is positioned securely into the groove.
Press TIMER and the display will show the countdown prior to
zeroing the blank. "SIP" will blink during zeroing.
Wait for a few seconds and the display will
show “-0.0-”. Now the meter is zeroed
and ready for measurement.
Remove the blank.
10 mL
# 2
x 12
# 1 # 2
x 12
# 1