59 Cyanuric Acid
CYANURIC ACIDCYANURIC ACID
CYANURIC ACIDCYANURIC ACID
CYANURIC ACID
SPECIFICATIONS
Range 0 to 80 mg/L
Resolution 1 mg/L
Accuracy ±1 mg/L ±15% of reading
Typical EMC ±1 mg/L
Deviation
Light Source Tungsten lamp with narrow band interference
filter @ 525 nm
Method Adaptation of the turbidimetric method. The reac-
tion between cyanuric acid and the reagent
causes a white suspension in the sample.
REQUIRED REAGENTS
Code Description Quantity
HI 93722-0 Powder reagent 1 packet
REAGENT SETS
HI 93722-01 Reagents for 100 tests
HI 93722-03 Reagents for 300 tests
For other accessories see page 141.
MEASUREMENT PROCEDURE
• Select the program number corresponding to Cyanuric Acid on the
secondary LCD by pressing PROGRAM W and V.
• Fill the cuvet with 10 mL of unreacted sample
(up to the mark) and replace the cap.
• Place the cuvet into the holder and
ensure that the notch on the cap is
positioned securely into the groove.
• Press ZERO and “SIP” will blink on the display.
• Wait for a few seconds and the display
will show “-0.0-”. Now the meter is
zeroed and ready for measurement.
10 mL