60
Cyanuric Acid
Fill a graduated beaker
up to the 25 mL mark
with the sample, add the
content of one packet of
HI 93722 reagent and
swirl gently to mix.
Fill a second cuvet with 10 mL of the
reacted sample up to the mark. Replace
the cap.
Reinsert the cuvet into the instrument.
Press TIMER and the display will show the countdown prior to the
measurement or, alternatively, wait for 45 seconds and press
READ DIRECT. In both cases "SIP" will blink during measurement.
The instrument directly displays concentration in mg/L of cyanuric
acid on the Liquid Crystal Display.