97
NICKEL HIGH RANGENICKEL HIGH RANGE
NICKEL HIGH RANGENICKEL HIGH RANGE
NICKEL HIGH RANGE
SPECIFICATIONS
Range 0.00 to 7.00 g/L
Resolution 0.01 g/L
Accuracy ±4% of reading
Typical EMC ±0.02 g/L
Deviation
Light Source Tungsten lamp with narrow band interference
filter @ 575 nm
Method Adaptation of the photometric method. The reaction
between nickel and the reagent causes a blue
tint in the sample.
REQUIRED REAGENTS
Code Description Quantity
HI 93726-0 Powder reagent 1 packet
REAGENT SETS
HI 93726-01 Reagents for 100 tests
HI 93726-03 Reagents for 300 tests
For other accessories see page 141.
MEASUREMENT PROCEDURE
• Select the program number corresponding to Nickel HR on the
secondary LCD by pressing PROGRAM W and V.
• Fill the cuvet up to the mark with 10 mL of
unreacted sample and replace the cap.
• Place the cuvet into the holder and
ensure that the notch on the cap is
positioned securely into the groove.
• Press ZERO and “SIP” will blink on the display.
• Wait for a few seconds and the display will
show “-0.0-”. Now the meter is zeroed
and ready for measurement.
10 mL
Nickel HR