96Molibdenum
Press TIMER and the display will show the countdown prior to the
measurement or, alternatively, wait for five minutes and press
READ DIRECT. In both cases, "SIP" will blink during measurement.
The instrument directly displays concentration in mg/L of molyb-
denum on the Liquid Crystal Display.
INTERFERENCES
Interference may be caused by:
Aluminum above 50 mg/L
Chromium above 1000 mg/L
Copper above 10 mg/L
Iron above 50 mg/L
Nickel above 50 mg/L
Nitrite, as NO
2
Sulfate above 200 mg/L
Highly buffered samples or with extreme pH may exceed the buffering
capacity of the reagents.