40
Total Chlorine HR
5 mL
TOTAL CHLORINE HIGH RANGETOTAL CHLORINE HIGH RANGE
TOTAL CHLORINE HIGH RANGETOTAL CHLORINE HIGH RANGE
TOTAL CHLORINE HIGH RANGE
SPECIFICATIONS
Range 0.0 to 10.0 mg/L
Resolution 0.1 mg/L
Accuracy ±0.1 mg/L ±3% of reading
Typical EMC ±0.1 mg/L
Deviation
Light Source Tungsten lamp with narrow band interference
filter @ 525 nm
Method Adaptation of the EPA DPD method 330.5. The
reaction between free chlorine and the DPD re-
agent causes a pink tint in the sample.
REQUIRED REAGENTS
Code Description Quantity
HI 93701-0 DPD 1 packet
HI 93734B-0 Free & Total Chlorine HR Reagent 5 mL
HI 93734C-0 Total Chlorine HR Reagent 3 drops
REAGENT SETS
HI 93734-01 Reagents for 100 tests
HI 93734-03 Reagents for 300 tests
For other accessories see page 141.
MEASUREMENT PROCEDURE
• Select the program number corresponding to Total Chlorine HR on
the secondary LCD by pressing PROGRAM W and V.
Add to the cuvet 5 mL of HI 93734B reagent
by means of the 5 mL syringe.