EasyManuals Logo

Hanna Instruments C 99 series User Manual

Hanna Instruments C 99 series
148 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #39 background imageLoading...
Page #39 background image
39
mL of unreacted sample. Replace the cap and shake gently
again.
Reinsert the cuvet into the instrument.
Press TIMER and the display will show the countdown prior to the
measurement or, alternatively, wait for 2 minutes and 30 seconds
and press READ DIRECT. In both cases “SIP” will blink during
measurement.
The instrument directly displays concentration in mg/L of total
chlorine on the Liquid Crystal Display.
Note: free and total chlorine have to be measured separately with
fresh unreacted samples following the related procedure if both
values are requested.
INTERFERENCES
Interference may be caused by: Bromine, Iodine, Ozone, Oxidized
forms of Chromium and Manganese.
In case of water with hardness greater than 500 mg/L CaCO
3
, shake
the sample for approximately 2 minutes after adding the powder
reagent.
In case of water with alkalinity greater than 250 mg/L CaCO
3
or
acidity greater than 150 mg/L CaCO
3
, the color of the sample may
develop only partially, or may rapidly fade. To resolve this, neutralize
the sample with diluted HCl or NaOH.
Total Chlorine

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Hanna Instruments C 99 series and is the answer not in the manual?

Hanna Instruments C 99 series Specifications

General IconGeneral
BrandHanna Instruments
ModelC 99 series
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals