98Nickel HR
Remove the cuvet and add the
content of one packet of HI
93726 reagent. Replace the cap
and shake gently until dissolu-
tion is complete.
Reinsert the cuvet into the instrument.
Press TIMER and the display will show the countdown prior to the
measurement or, alternatively, wait for 1 minute and press READ
DIRECT. In both cases "SIP" will blink during measurement.
The instrument directly displays concentration in g/L of nickel on
the Liquid Crystal Display.
INTERFERENCES
Interference may be caused by copper.