Ammonia HR
Press ZERO and “SIP” will blink on the display.
Wait for a few seconds and the display will
show “-0.0-”. Now the meter is zeroed
and ready for measurement.
Remove the cuvet.
Add 4 drops of HI 93733A
Nessler Reagent. Replace the
cap and mix the solution.
Reinsert the cuvet into the instrument.
Press TIMER and the display will show the countdown prior to
measurement or, alternatively, wait for 3 minutes and 30 seconds
and then press READ DIRECT. In both cases the display will show
“SIP” during measurement.
The instrument directly displays concentration in mg/L of Ammo-
nium ion (NH
4
+
) on the Liquid Crystal Display.
To convert the reading to mg/L of ammonia (NH
3
), multiply by a
factor of 0.944.
To convert the reading to ammonia nitrogen (NH
3
-N), multiply by
a factor of 0.776.
INTERFERENCES
Interference may be caused by:
acetone, alcohols, aldehydes, glycine, hardness above 1 g/L, iron,
organic chloramines, sulfide, various aliphatic and aromatic amines.