EasyManua.ls Logo

Hanna Instruments C 99 series - COD MR Interferences

Hanna Instruments C 99 series
148 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
113
Leave the vials in the tube rack to cool to room temperature. Do not
shake or invert them anymore otherwise the samples may become
turbid.
Select the program number corresponding to Oxygen Demand,
Chemical (COD) - MR on the secondary LCD by pressing PROGRAM
W and V.
Place carefully the COD-vial adap-
tor in the instrument and check
that it fits properly.
Place the blank into the holder.
Press ZERO and “SIP” will blink on the display.
Wait for a few seconds and the
display will show “-0.0-”. Now
the meter is zeroed and ready
for measurement.
Remove the blank.
Place the sample into the holder.
Press READ and “SIP” will blink during measurement.
The instrument directly displays concentration in mg/L of oxygen
demand on the Liquid Crystal Display.
INTERFERENCES:
chlorides > 2000 ppm
COD MR

Table of Contents

Related product manuals