Add some more sample, to fill the bottle completely. Replace the
cap again and ensure that a part of the sample spills over. This
is to make sure that no air bubbles have been trapped inside,
which would corrupt the reading.
Invert several times the bottle. The
sample becomes orange-yellow and
a flocculant agent will appear.
Let the sample stand and the flocculant agent
will start to settle.
After approximately 2 minutes, when the upper
half of the bottle becomes limpid, add 10 drops
of HI 93732C.
Replace the cap and invert the bottle until dissolution of the
settled flocculant is complete. The sample is ready for measure-
ment when it is yellow and completely limpid.
Fill the cuvet up to the mark with 10 mL of the
unreacted (original) sample, and replace the
cap. This is the blank.
Place the cuvet into the holder and
ensure that the notch on the cap is
positioned securely into the groove.
Press ZERO and “SIP” will blink on the display.
10 mL
x 10
Dissolved Oxygen

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Hanna Instruments C 99 series and is the answer not in the manual?

Hanna Instruments C 99 series Specifications

General IconGeneral
BrandHanna Instruments
ModelC 99 series
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals