51 Color of Water
# 3
Fill another cuvet up to the mark with
unfiltered sample and replace the cap.
This is the apparent color.
Filter 10 mL of sample through a filter with a
0.45 µm membrane into the third cuvet, up
to the 10 mL mark and replace the cap. This is
the true color.
Insert the apparent color cuvet (# 2)
into the instrument and ensure that
the notch on the cap is positioned
securely into the groove.
Press READ DIRECT and "SIP" will blink on the display.
The meter directly displays the value of apparent color in PCU on
the Liquid Crystal Display.
Remove the cuvet, insert the true color
cuvet (# 3) into the instrument and
ensure that the notch on the cap is
positioned securely into the groove.
Press READ DIRECT and "SIP" will blink on the display.
The meter directly displays the value of true color in PCU on the
Liquid Crystal Display.
10 mL
# 2
# 2
# 3