EasyManuals Logo

Hanna Instruments C 99 series User Manual

Hanna Instruments C 99 series
148 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #71 background imageLoading...
Page #71 background image
71
Total Hardness HR
Add 2 drops of HI 93735B Buffer reagent.
Replace the cap, shake
gently to mix and wait
for 15 seconds.
Place the cuvet into the holder and
ensure that the notch on the cap is
positioned securely into the groove.
Press ZERO and "SIP" will blink on the display.
Wait for a few seconds and the display will
show "-0.0-". Now the meter is zeroed and
ready for measurement.
Remove the cuvet and add the content of 1 packet
of HI 93735C Fixing reagent.
Replace the cap and
shake gently till
dissolution is complete.
Place the cuvet into the holder and
ensure that the notch on the cap is
positioned securely into the groove.
x 2