EasyManua.ls Logo

Hanna Instruments C 99 series - Page 87

Hanna Instruments C 99 series
148 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
87
Place the cuvet into the holder and
ensure that the notch on the cap is
positioned securely into the groove.
Press ZERO and "SIP" will blink on the display.
Wait for a few seconds and the display
will show "-0.0-". Now the meter is
zeroed and ready for measurement.
Remove the cuvet.
Fill another graduated mixing cylin-
der up to the 25 mL mark with the
sample.
Add the content of one packet
of HI 93746 TPTZ reagent,
close the cylinder and shake
vigorously for 30 seconds. This
is the reacted sample.
Fill a cuvet with 10 mL of the reacted
sample up to the mark and replace the
cap.
25 ml
10 mL
12
3
6
9
30"
Iron LR

Table of Contents

Related product manuals