111  HI520 Events Management System
Warning Description
Continue Logging
Stop Control
Alarm Relay &LED
Status LED
Hold Relay & LED
WARNING
_
CH1
_
ODO
_
CAP
_
TIMEOUT
HI7640‑58
Optical DO only
Expired Smart Cap.
Replace the Cap.
NO NO
Off
CC CC
Off
CC
WARNING
_
CH1
_
ODO
_
CAP
_
TIMEOUT
_
SOON
Smart Cap expires soon.
Cap is due for replacement.
NO NO
Off
CC CC
Off
CC
WARNING
_
RTC
_
SET
_
TO
_
FIRST
_
VALUE
RTC is running. NO NO
Off
CC CC
Off
CC
WARNING
_
LOT
_
LOG
_
FULL
Current lot log file has reached 8600
records. A new file will be generated
automatically and old data deleted.
Save current data on USB and delete
the file to prevent data loss.
NO NO
Off
CC CC
Off
CC
WARNING
_
LOT
_
LOG
_
MAX
_
INDEX
_
ASSIGNED
Maximum number of logged files per
day has been reached (100).
Save current data on USB and delete
the file to prevent data loss.
NO NO
Off
CC CC
Off
CC
WARNING_HIGH_MATH_CHANNEL
Primary parameter exceeds math
channel High set value. Mask time has
not yet expired to generate an alarm.
Check tanks and all installed devices are
functioning correctly.
NO NO
Off
CC CC
Off
CC
WARNING_LOW_MATH_CHANNEL
Secondary parameter is below secondary
parameter math channel High set value.
Mask time has not yet expired to
generate an alarm.
NO NO
Off
CC CC
Off
CC
WARNING
_
CONTROL
_
DELAY
Start up is delayed and control is not
running.
NO NO
Off
CC CC
Off
CC
WARNING
_
MAIN
_
OUTOFF
_
COMPRANGE
Main parameter is outside
compensation range.
NO NO
Off
CC CC
Off
CC
WARNING
_
TEMP
_
OUTOFF
_
COMPRANGE
Temperature outside compensation
range
NO NO
Off
CC CC
Off
CC
WARNING
_
TEMP
_
SENSOR
_
BROKEN
Temperature sensor not working.
Replace the probe.
NO NO
Off
CC CC
Off
CC