Technical Menu 60
12.2. SIMPLE CLEANING
• With item selected, press Set to modify.
• Use the
ud keys to modify the flashing digit.
• Press CFM to save.
Item Option Screenshot
Rinse Time 5 to 300 seconds
Recovery Time 5 to 120 seconds
Rinse Relay Displays allocated rinse relay
Note: Controller validates configured Setup when attempting to exit Menu and directs the user to any invalid
parameters. At prompt to save changes, press YES.
13. TECHNICAL MENU
The technical menu is used for on‑site, single point pressure calibration (Press. Calibration), Analog Output
calibration (AO Calibration) and firmware update.
Current pressure values are entered manually and reading is displayed in mmHg.
Enable
b
Setup
Navigation:
• Press the
m key from the Measure mode.
• With Technical Menu selected, press Setup to enter the screen.
• Use the
ud keys to navigate options.
• With option highlighted, press the displayed functional key to start calibration or firmware update.