63 Manual Mode
14. MANUAL MODE
When Manual Mode is selected, Setup is visible. Select Setup to open up Relays (with their configured function)
and Analog Outputs submenu structure.
Enable
b
Setup
When relays are turned to on, it can manually test the relay connection and operation (relay contact opening
and closing) and also the operation of the associated equipment, and is a useful feature to prime a dosing
pump for example. The current loop(s) can be tested by setting a current value and verifying it at the outputs.
Navigation
• From Main Menu, press the
ud keys to select Manual mode.
• With option selected, press Setup to enter the screen.
• Press the
ud keys to move between the five relays and two or four analog outputs.
Relay Alarm
Option: On, Off
Relay set to be On, keeps its status for maximum 60 minutes before it switches Off; or user leaves Manual Mode.
Relay x
Option: On, Off
Relay set to be On, keeps its status for maximum 60 minutes before it switches Off; or user leaves Manual Mode.
Analog Output AO x
Option: 0.0 to 22.0 mA
1. In Manual Control screen, press the
ud keys to move to AO x.
2. With AO x selected, press Set, to modify. Use the
ud keys to modify the flashing digit.
3.
Press CFM, to save. The analog remains at the current set for 60 min. until it resumes previous current value.