137  Measuring with EC Probes
Calibration Messages & Warnings
Message & Description Recommended Action
Wrong Std”
The reading exceeds the expected
value. Calibration can not be
confirmed.
Check that correct calibration solution
has been used and / or clean the
probe.
“Temperature Error
The temperature of the solution is out
of temperature compensation interval.
Use fresh calibration solution and / or
clean the temperature sensor.
“Maximum window
During a process calibration, the
calibration value exceeds upper
boundary value.
Change calibration point value, clear
calibration or exit by pressing the back
key.
22.5. MAINTENANCE
Clean the probe regularly to prevent debris buildup between rings or blockage of the vent hole (four‑ring
probes).
Rinse the probe thoroughly as water residue may not be visible.
Inspect all sensor connectors for corrosion and replace if necessary.
Cleaning
Dirty or improperly cleaned probes can result in erratic and inaccurate readings.
Remove and inspect the probe during scheduled service intervals.
Dry clean the sensor with a soft bristle brush to loosen any debris.
For a more thorough cleaning:
Use a cloth and warm water with a soapy surfactant to clean and follow with a thorough rinse with purified
(deionized) water. Ensure that the holes and cell channel in the sensor are free of foreign material.
Flush with purified water after cleaning.
Four‑ring probe cleaning
Remove and inspect the probe during scheduled service intervals.
Clean off the external sheath with a soft cloth and surfactant solution.
Rinse the probe under a stream of running tap water to remove salt or minerals.
Jet the tap water stream through the opening to dislodge any debris.
If strictly necessary, carefully remove the outer plastic sheath to disassemble the probe.
Clean with a warm water/surfactant mixture and follow with a through rinsing with purified water.
Allow pieces to dry and reassemble.
Storage
EC probes should always be stored dry after cleaning in distilled water.
After long‑term storage or cleaning, calibration is required.