149  Measuring with Optical Dissolved Oxygen (DO) Probes
24.2. GENERAL INSTALLATION CONSIDERATIONS
Probes are easily installed using the ¾” NPT external threads.
Hand tighten the probe in position. Then, depending on the process, tighten one or two turns with a wrench
to secure in place. Do not exceed the 10 Nm (7.3 lb‑ft) torque specification for the probe sensor.
Protect the probe and the Smart Cap from strong flow to prevent unstable readings. In turbulent aeration
basin installations, place the probe in a weir for more accurate readings.
Position the probe so that it does not trap air bubbles at the Smart Cap.
Deposits of foreign material should not be allowed to accumulate within the sensing area.
Protect the sensing surface from blunt objects and is keep it clean.
Provisions must be made for the removal of the probe from the process.
Consider probe accessibility for maintenance when selecting placement.
Note: See Optical DO industrial probes manual (MAN7640-58) for series configuration options and detailed
specifications.
24.3. INSTALLATION SCHEMES & MOUNTING ACCESSORIES
Accessories are sold separately!
24.3.1. Probe Dimensions
3/4’’ NPT
Ø 17 mm (0.7”)
Ø 28 mm (1.1”)
3/4’’ NPT
172 mm (6.7”)
64 mm (2.5”)
43 mm (1.7”)
Figure 71: HI7640‑58ZZ optical DO with attached cable
172 mm (6.7”)
Figure 72: HI7640‑5800 optical DO with DIN connector
24.3.2. Probe Connection
Align the pins and key then push the plug into the socket.
Rotate the collar to lock in place.
Note: Probe connection (probe with integral DIN connector) and probe wiring (probe with attached cable)
must be carried out with the controller disconnected from power.