Math Channel  48
8. MATH CHANNEL
Match channel dispalys the mathematical result of the measurement between two probes with identical
measurement configuration and unit.
Channel can only be configured when:
Both probes are connected and initialized
Both probes are of the same type (pH, ORP, EC, DO) have same Measure Mode, and Measurement Unit
Notes: Channel data is not logged in the data Log files.
Channel alarms set neither CH1 control nor CH2 control on Hold, when enabled and activated.
Channel & function availability by parameter
Function pH ORP
EC
DO OPDO
EC TDS RES Sal% Sal ppt Sal psu
Difference
Ratio
% Passage
% Rejection
8.1. MATH CHANNEL SETUP
Navigation
From Menu, use
ud keys to select Math Channel.
Use Math Channel submenu to enable math function.
Math functions
Function Formula Range
Default
Alarm Low
Default
Alarm High
Difference (a – b) PHL to +PHL PHA +PHA
Ratio (a / b) 0.0 to 500.0 0.0 100.0
% Passage (a / b) × 100 0.0 % to 500.0 % 0.0 % 100.0 %
% Rejection [1 – (a / b) ] × 100 – 500.0 % to +500.0 % 0.0 % 100.0 %
a = main Measurement (CH1)
b = main Measurement (CH2)
PHL =
Probe High Limit PHA = Probe High Alarm
Low and High alarms can be set on the channel.