51  Outputs
10.2. ANALOG OUTPUTS
Note: Controller model determines the number of relays and analogs.
Navigation
From Analog Outputs, press Setup.
Press the
ud keys to navigate between parameters.
Enable
b
Setup
Enable
b
Setup
ud
Press the b key to return to the menu without saving.
At prompt, enter the passcode.
At prompt, with the password enabled, select YES to place unit in HOLD and start modifying parameters
Note: We suggest users make configuration changes from the beginning of this Menu structure going forward,
because of the menu references parameters that were set earlier in the submenu.
Mode
Option: Disabled, Track Channel
With Mode selected, press Modify to toggle options.
Disabled indicates that analog output has not been allocated to any function .
With Track Channel selected the analog output follows a specific parameter.
Data Channel
Option: CH1, CH2, Math channel