Measuring with Galvanic Dissolved Oxygen (DO) Probes 142
23.3.6. Installation tee, user supplied
• Orient the probe with the sensor facing the flow.
• Wrap PTFE tape around the probe’s top threads and fittings before mounting.
FLOW
FLOW
2
1
3
23.4. CONFIGURABLE MEASUREMENT PARAMETERS
Temperature Source (Temp. Source)
Option: Probe, Manual
• With item selected, press Modify for the drop‑down options list.
• Use the
ud keys to scroll between Probe or Manual.
• Press Select to save.
Manual Temperature (Man. Temp.) Value
Default value is 25 °C
• With item selected, press Set to modify the value.
• Use the
ud keys to modify the flashing value.
• Press CFM, to save.
Temp. Source has to be set as Manual.
1 Tee
2 Adapter
3 Probe
Mounting accessories
Tee fitting size Adapter size
1” from 1” to ¾” NPT
1 1/4” from 1 1/4” to ¾” NPT
1 1/2” from 1 1/2” to ¾” NPT
2” from 2” to ¾” NPT