143  Measuring with Galvanic Dissolved Oxygen (DO) Probes
Salinity Factor[g/l]
Option: 0 to 70 g/L
The salinity correction factor is the ratio of the solubility of oxygen in water at a particular salinity to its solubility
in fresh water at an identically specified water temperature and barometric pressure.
With item selected press Set.
Use the
ud keys to modify the flashing value.
• Press CFM to save the value.
Measurement Mode (Meas.Mode)
Option: DO_%Sat, DO_Conc
Allows users to select between measuring DO saturation and DO concentration.
With item selected, press Modify for the drop‑down options list.
Use the
ud keys to scroll between options.
• Press Select to save.
Measurement Unit (Meas.Mode)
Option: mg/L or ppm (with Meas.Mode set to DO_Conc)
Note: With Meas.Mode set to DO_%Sat, measurements are displayed in %Sat.
Averaging Samples
Option: 1 to 60
Average sampling is a software filter to minimize sensor noise and provide more stable readings. Allows users
to get a representative reading of the “average” value from flowing water.
Averaging affects measurement. If a fast response is needed, this value should be kept low.
With item selected, press Set.
Use the
ud keys to modify the number of samples to average.