6.2. PROBE SETTINGS, GENERAL PARAMETERS
This section groups configurable Probe Settings items common to all wired probes regardless of measured
parameter as well as probe information details.
Calibration timeout, Temp. source, Manual Temperature Value, setting the temperature Offset value
follow the same steps regardless of wired parameter.
pH screes only are referenced as examples here. Information screens are probe specific.
For parameter‑specific information on configurable options please refer to the following sections:
` 21 Measuring with pH & ORP Probes
` 22 Measuring with EC Probes
` 23 Measuring with Galvanic Dissolved Oxygen (DO) Probes
` 24 Measuring with Optical Dissolved Oxygen (DO) Probes
Probe Info
Option: Model, Firmware, Serial No.(number), Factory cal.(calibration)
Example of probe information screen with wired pH probe.
Calibration Timeout (Cal. Timeout)
Option: Disabled, 1 to 99 days
With Cal. Timeout selected, press Set, to modify.
Press the
ud keys to modify the flashing value and press CFM to save.
Cal. Timeout is used to send a reminder to recalibrate the probe. "
а
" is displayed in the Title & Status area.
Press the h key to access calibration message.
Temp. Offset & Temperature Calibration Procedure
Option: –5.0 to 5.0 °C, –9.0 to 9.0 °F
With Temp. Offset selected, press Set.
Press the
ud keys to modify the flashing value and press CFM to save.
A positive value adds to the displayed temperature.
A negative value decreases the displayed temperature value.
To obtain the temperature offset, see step 3, Temperature Calibration Steps.