Inputs 54
Fixed Value
• With Fixed Value selected, press Set. The value will flash indicating it can be modified.
• Press the
ud keys to increase or decrease the value. Press CFM to save the value.
• Press the
b key to return to the menu.
Out 22 mA ‑ OnALARM
Option: Enabled, Disabled
With Out 22 mA ‑On ALARM selected press the corresponding virtual key to enable or disable function.
When enabled, it drives the analog output to 22 mA in an alarm condition.
Note: The controller validates the configured Setup when attempting to exit Menu and directs the user to any
invalid parameters. At prompt to save changes, press YES.
11. INPUTS
Note: Setup selections do not change if a new parameter probe is used on controller.
With Inputs selected press Setup to open a submenu structure for Input 1 and Input 2.
Enable
b
Setup
u
d
Both inputs are configured the same way. Verify the wiring before configuration.
Navigation
• From Main Menu, use the
ud keys to select Inputs.
• Use the
ud keys to toggle between options.
• With option selected, press Setup, to open the selected input.
If required:
• At prompt, enter the passcode.
• At prompt, press YES to place unit in HOLD.
Each of the two inputs can be configured as disabled or used to trigger Hold Mode or a Cleaning cycle from a
remote trigger switch.
The active level of the input can be set High or Low.