53  Outputs
Output Range
Option: 0 ‑20 mA, 4 ‑20 mA
With Output Range selected, press the corresponding virtual key, to toggle mA output range: 0 ‑20 mA or 4 ‑20 mA.
0 mA or 4 mA Value
Option: measured parameter, CtrlSetP1 or CtrlSetP2
With 0 mA (or 4 mA) Value selected, press Set.
The value will flash indicating it can be modified.
Press the
ud keys to increase or decrease the value.
• Press CFM to save.
20 mA Value
Option: measured parameter, CtrlSetP1 or CtrlSetP2
With 20 mA Value selected, press Set. The value will flash indicating it can be modified.
Press the
ud keys to increase or decrease the value.
• Press CFM to save.
HOLD Value
Option: Last Frozen, Fixed Value
With HOLD Value selected, use the virtual key to toggle between FixedValue or LastFrozen.
Last Frozen indicates output being held at present level, prior to hold.
Fixed Value indicates output being driven to a configured value during hold.
Note: Value is set in the next parameter; Fixed Value.