127 Measuring with EC Probes
216 mm (8.5”)
Figure 67: HI7630‑4800 four‑ring EC probe with DIN connector
22.2.2. Probe Connection
Align the pins and key then push the plug into the socket.
Rotate the collar to lock in place.
Note: Probe connection (probe with integral DIN connector) and probe wiring (probe with attached cable)
must be carried out with the controller disconnected from power.
22.2.3. In‑line Mounting with Probe Saddle
54
2
31
1 Pipe
2 Sensor tip
3 Saddle
4 Probe fitting kit
5 Probe
Note: To prevent air entrapment inside the electrode the probe can be rotated and horizontal mounting is the
recommended option.
Required accessories
Pipe size Saddle code HI7630‑28zz HI7630‑48zz
Ø 50 mm (2”) BL120‑550
Ø 63 mm (2½”) BL120‑563
Ø 75 mm (3”) BL120‑575
22.2.4. Installation Tee
• Install the probe in a tee, horizontally, and direct the flow into the opening at the bottom of the probe.
• Upward flow direction must always be ensured.
• Maintain a stable flow rate to minimize interferences from bubbles and settling of particulates.
• Wrap PTFE tape around the probe’s threads and fittings before mounting.