131  Measuring with EC Probes
• Press Select to save.
Manual Temperature Value (Man. Temp. Value)
Default value is 25 °C.
With item selected, press Set to modify the value.
Use the
ud keys to modify the flashing value, down to minimum or up to maximum probe limits
(‑20 to 120 °C/68 to 248 °F).
• Press CFM to save.
This temperature is used when Temp. Source is set to Manual.
Reference Temperature (Ref.Temp.[°C])
Option: 15.0 °C to 25.0 °C
This value is used for temperature compensated conductivity. All EC measurements will be referenced to the
conductivity of a sample at this temperature.
With item selected, press Set to modify.
The flashing value indicates that it can be modified by using the
ud keys.
Press CFM to save.
Temperature Coefficient (Temp.Coef[%/°C])
Option: 0.00 to 10.00 %/ °C
Temperature coefficient is a function of the solution being measured. For freshwater samples, the temperature
coefficient is approximately 1.90 %/ °C.
With item selected, press Set to modify. The flashing value indicates that it can be modified by using the
ud keys.
• Press CFM to save.