Measuring with Optical Dissolved Oxygen (DO) Probes  156
24.5.3. Clear Calibration
1. Press CAL to enter calibration mode.
2. CLR option is displayed for a few seconds.
3. Press CLR to clear previous calibration.
4. Press YES to confirm deletion. “Calibration Erased” message screen is displayed for a few seconds then
the controller returns to user calibration mode.
5. Press the
b (back) key to return to the menu.
Calibration Messages & Warnings
Message & Description Recommended Action
Wrong Std”
The reading exceeds the expected
value. Calibration can not be
confirmed.
Check that correct calibration
solution has been used and / or
clean the probe.
“Temperature Error
The temperature of the solution is out
of temperature compensation interval.
Use fresh calibration solution and / or
clean the temperature sensor.
MAINTENANCE
General Maintenance
Inspect O‑ring for nicks or other damage. Replacing the o ring is advised.
Do not substitute other grease or lubricants as it may cause the O‑ring to swell.
After long‑term storage or cleaning, calibrate the probe.
After use, rinse the probe with tap water and dry it.
The DO cap must be kept hydrated.
Cleaning the Smart Cap
Use a mild detergent and a soft‑bristled toothbrush to clean.
Rinse with water after cleaning and dry with a laboratory tissue.
Hydrate in purified water before use.
Note: Smart Caps need to be replaced every year.