37  Channel
Parameter
Option: see 2.2 Supported Probe Series Configurations for supported parameters
Parameter, measurement units, minimum and maximum probe values, Hysteresis, Deviation, Dead Band values,
Control Period, Reset Time, Rate Time, Dead Band Gain depend on configured Control Mode option (i.e. ON/
OFF, Proportional, PID) in Probe Settings submenu.
With Parameter selected, press the virtual key, and press the
b
key to save or
d
key to move to next
parameter.
Enabled
Set point option has to be enabled to set up the Control function.
With Enabled selected, press Enable or Disable. A check mark will appear to confirm selection.
Press the
b key, to exit or d key to move to next parameter.
Value
This parameter defines the Setpoint value. Setpoint must be enabled first
With Value selected, press Set.
Press the
ud keys to edit the required value within minimum / maximum probe limits (e.g. 0.00
to 12.00 pH), displayed blinking.
• Press CFM, to save.
Note: A Control Setpoint value cannot be the same as the Alarm Setpoint.
Mode (Setpoint must be enabled first)
Option: ON/OFF, Prop., PID
The Mode parameter defines the type of control the controller will use i.e.ON/OFF, Prop., or PID.
The Setup virtual key is used after selecting Mode, to set additional settings.
See 18 Control Modes & Algorithms for detailed information.